Single-use Chromatography Benchtop - 3 LPM
The brand-agnostic chromatography skid gives you unmatched options for filters, columns, and other components—and even the control system.
Maximum flexibility
- Customize every aspect of the skid to meet your unique needs—whether it’s removing inlets/outlets, analytics, filters, columns, we can handle any combination of features/functions
- Column- and resin-agnostic—allows you to choose your preferred brand of columns/filters—whether Sartorius, Millipore, GE, Repligen, or a custom pre-filter
- Flexible automation based on ISA 88 Standards for batch operations: run manually, load individual steps (Prime, Wash, Load, Elution, etc.), or run a fully automated sequence
- Run process in upflow or downflow
- Transition from flow control to pressure control if needed (in case of filter fouling)
- Run, bind and elute, or flow-through process steps
- Add multiple column connections for serial operation, if needed
Optimized performance and fit
The chromatography skid is engineered to accommodate any low flow rate requirements. The result: column sizes and resin quantities can be optimized for the specific process needs, keeping raw material costs as low as possible. A broad dynamic range enables optimization of skid size for every application. And, the combination of optimally sized inline analytical sensors and well-less measurement devices reduce hold up.
Single-use technology for a safe and sterile process
- Fully sterilized, single-use tubing set, complete with integrated sensors, minimizes the risk of contamination
- Sterile-to-sterile connections for inlets/outlets reduce contamination risks
- Uses only the highest quality components, delivering USP <88> Class VI certified single-piece tubing sets that ensure sterility throughout use
- Batch fed or continuous
- Flexible automation
- Brand-agnostic filters, columns, and other components
- Adapts to your process requirements
- Cost-effective scalability
- Safe and sterile single-use technology
- A choice of models
Attribute | C125 | C250 |
---|---|---|
Tubing ID | 0.125 in | 0.25 in |
Flow Rates | 1-500 mL/min | 16-3000 mL/min |
Pressure | 0-60 psi/4 bar | 0-60 psi/4 bar |
pH | 2-12 pH | 2-12 pH |
Conductivity | 0-100 mS/cm | 0-100 mS/cm |
UV Wavelength | 0-3 AU Dual Wavelength-280 nm and 300 nm standard. Customer configurable to any 2 wavelengths between 255-1000 nm | 0-3 AU Dual Wavelength-280 nm and 300 nm standard. Customer configurable to any 2 wavelengths between 255-1000 nm |
Temperature Range | 2-50°C | 2-50°C |
Humidity (non-condensing) | 10-90% | 10-90% |
Utility Required, Skid Power* | 110-120 VAC | 110-120 VAC |
Wetted Materials | USP <88> Class VI and animal free | USP <88> Class VI and animal free |
Inlets | 8 (4 ea inlet header) | 8 (4 ea inlet header) |
Outlets | 2 | 2 |
Flow Accuracy | ±3% | ±3% |
Hold-Up Volume-Inlet to Column | 50 mL | 95 mL |
Hold-Up Volume-Column to Outlet | 10 mL | 36 mL |
Dims (W x D x H)** | 610 x 610 x 610 mm | 610 x 610 x 610 mm |
Sensor Performance | ||
Pressure | 0-6 psi (± 2%) 6-30 psi (± 3%) 30-60 psi (± 5%) | 0-6 psi (± 2%) 6-30 psi (± 3%) 30-60 psi (± 5%) |
Conductivity | 0-100 mS/cm (± 5%) | 0-100 mS/cm (± 5%) |
Flow | 0-4 LPM (± 3%) | 0-8 LPM (± 3%) |
pH | 2-12 pH (± 0.1) | 2-12 pH (± 0.1) |
UV | 0-2 AU (± 1%) 2-3 AU (± 2%) | 0-2 AU (± 1%) 2-3 AU (± 2%) |
Related Products
At Liquidyne, we prioritize our commitment to quality and consistency. In 2017 we achieved ISO 9001:2015 Quality Management System certification to ensure we can repeatedly deliver consistently high-quality products and solutions. Being ISO 9001:2015 QMS Certified by Orion Registrar, Inc, together with our USP Class VI Testing, ASME Bioprocessing Equipment (BPE), and EHEDG certification guarantee, we have the methodologies to assure quality. We will quickly and accurately rectify any issues if any of our products are ever in non-conformance.